- BPM-1000(MATRON)
Matron
It is ALL in your handgrip, NIBP, IBP, SpO2, ECG, Respiration, Temperature, Battery operation and Stress-free data scrolling functions
but a few...
Take it to the Limit.
Protable design and features with an easy-grip handle.
Battery operation keeps you mobile anytime, any where.
¡Å Details
- Display
- Screen : 10.4" color TFT LCD
- Resolution : 640 x 480
- Trace : 6 Max.
- Sweep Speed " 603 /12.5 / 25 / 50mm/s
- Indicator : AC power LED, Battery charging LED
- Alarm suspend LED, Alarm LED
ECG
- Input : 5-lead (I, II, III, aVL, aVR, aVF, V)
- ECG waveform : 1Channer (the 2nd waveform cascade)
- 2Channer (Optional)
- Sweep speed : 6.3 / 12.5 / 25 / 50mm/s
- Gain (Sensitivity) : 5 / 10 / 15mm/mV
- Pulse range : 20 to 300bpm
- Pulse accuracy : ¡¾2bpm
- Lead-off Detection : Yes
- Arrhythmia Analysis : Yes
Respiration
- Measurement range : 0~150bpm
- Resolution : 1rpm
- Accuracy : ¡¾2rpm
- Apnea delay : 10~40 sec (in 5 sec. Steps)
NIBP
- Technique : Oscilometric
- Operation mode : manual
- Automatic (2/5/10/15/30/60/120min)
- Inflation pressure range : 0-300mmHg
- Inflation pressure Max : Adult : 300 ¡¾5mmHg
- Neonate : 150 ¡¾5mmHg
- Resoluction : 1mmHg
- Over-pressure protection : Yes
- Adult : Systolic(60-270), Diastolic(40-245), Mean(45-255)
- Neonate : Systolic(50-130), Diastolic(30-100), Mean(40-120)
SpO£²
- Measurement range : 20-100%
- Measurement range accuracy : ¡¾2%
- Pulse rate range : 20-300bpm
- Pulse rate accuracy : ¡¾2bpm
- SpO£² accuracy : ¡¾2% (70-100%), ¡¾3% (50-69%), 0-49% Unspecified
- Resolution : 1%
IBP (2 IBP / optional)
- Measurement range : -40 to 360mmHg
- Channel : 2 channel
- Resolution : ¡¾1mmHg
- Accuracy : ¡¾2mmHg
- Auto zero balancing range : ¡¾20mmHg
- Alarm Range : -40 to 360mmHg
Temperature(2 Temperature)
- Measurement range : 10~45¡É
- Accuracy : 0.1¡É
EtCO2 (optional)
- Measurement mode : Microstream
- Range :0~99mmHg
- Accuracy : 0~38mmHg ¡¾2mmHg
39~99mmHg ¡¾5%+0,08%
- for every 1mmHg above 38mmHg
- Respiration rate : 0~150 breaths per minute
- Respiration accuracy : ¡¾1breaths per minute from 0~70bpm
¡¾2bpm from 71~120bpm
¡¾3% from 12~150bpm
- CO£² waveform
- Recognition : 5mmHg ¡¾2mmHg bandwidth
- Start-up time : 30seconds typical
- Sampling rate : 50ml/minute ¡¾7.5ml/min
- Auto zero : <15seconds at 5, 10, 15 and 60minute then every 60minutes thereafter.