Human Health Care Service ¹àÀº ¿ôÀ½°ú ÇູÇÑ °¡Á·, »çȸ¸¦ ÀÌ·ç¾î °¡°Ú½À´Ï´Ù.

Äü¸Þ´º

»ó´ãÀüÈ­ Á¦Ç°±¸¸Å 033 434-9041~2 °í°´»ó´ã 033 434-9041~2

> Product > PRODUCT GROUP1

BPM-1000(MATRON)

Matron

It is ALL in your handgrip, NIBP, IBP, SpO2, ECG, Respiration, Temperature, Battery operation and Stress-free data scrolling functions
but a few...

Take it to the Limit.
Protable design and features with an easy-grip handle.
Battery operation keeps you mobile anytime, any where.

¡Å Details

  • Display
    - Screen : 10.4" color TFT LCD
    - Resolution : 640 x 480
    - Trace : 6 Max.
    - Sweep Speed " 603 /12.5 / 25 / 50mm/s
    - Indicator : AC power LED, Battery charging LED
    - Alarm suspend LED, Alarm LED

    ECG
    - Input : 5-lead (I, II, III, aVL, aVR, aVF, V)
    - ECG waveform : 1Channer (the 2nd waveform cascade)
    - 2Channer (Optional)
    - Sweep speed : 6.3 / 12.5 / 25 / 50mm/s
    - Gain (Sensitivity) : 5 / 10 / 15mm/mV
    - Pulse range : 20 to 300bpm
    - Pulse accuracy : ¡¾2bpm
    - Lead-off Detection : Yes
    - Arrhythmia Analysis : Yes

    Respiration
    - Measurement range : 0~150bpm
    - Resolution : 1rpm
    - Accuracy : ¡¾2rpm
    - Apnea delay : 10~40 sec (in 5 sec. Steps)

    NIBP
    - Technique : Oscilometric
    - Operation mode : manual
    - Automatic (2/5/10/15/30/60/120min)
    - Inflation pressure range : 0-300mmHg
    - Inflation pressure Max : Adult : 300 ¡¾5mmHg
    - Neonate : 150 ¡¾5mmHg
    - Resoluction : 1mmHg
    - Over-pressure protection : Yes
    - Adult : Systolic(60-270), Diastolic(40-245), Mean(45-255)
    - Neonate : Systolic(50-130), Diastolic(30-100), Mean(40-120)

    SpO£²
    - Measurement range : 20-100%
    - Measurement range accuracy : ¡¾2%
    - Pulse rate range : 20-300bpm
    - Pulse rate accuracy : ¡¾2bpm
    - SpO£² accuracy : ¡¾2% (70-100%), ¡¾3% (50-69%), 0-49% Unspecified
    - Resolution : 1%

    IBP (2 IBP / optional)
    - Measurement range : -40 to 360mmHg
    - Channel : 2 channel
    - Resolution : ¡¾1mmHg
    - Accuracy : ¡¾2mmHg
    - Auto zero balancing range : ¡¾20mmHg
    - Alarm Range : -40 to 360mmHg

    Temperature(2 Temperature)
    - Measurement range : 10~45¡É
    - Accuracy : 0.1¡É

    EtCO2 (optional)
    - Measurement mode : Microstream
    - Range :0~99mmHg
    - Accuracy : 0~38mmHg ¡¾2mmHg
                      39~99mmHg ¡¾5%+0,08%
    - for every 1mmHg above 38mmHg
    - Respiration rate : 0~150 breaths per minute
    - Respiration accuracy : ¡¾1breaths per minute from 0~70bpm
                                       ¡¾2bpm from 71~120bpm
                                       ¡¾3% from 12~150bpm
    - CO£² waveform
    - Recognition : 5mmHg ¡¾2mmHg bandwidth
    - Start-up time : 30seconds typical
    - Sampling rate : 50ml/minute ¡¾7.5ml/min
    - Auto zero : <15seconds at 5, 10, 15 and 60minute then every 60minutes thereafter.

¡Å ACC